A.8 F
UNCTION
#7
–
SPECIAL CONTROL FUNCTIONS
#7 function enables the user to perform special control functions.
Some of them should be used with
the extreme care.
#7 function
formats are defined as follows:
#7,AC;
This function returns auto calibration in the format
#7,AC,x;
#7,AC,x;
This function enables ( x = 1 ) or disables ( x = 0 ) the auto calibration and returns the following sequence
of characters:
#7,AC;
#7,AS;
Get settings for the Auto-Run function.
Response format:
#7,AS,e,HH,MM,hh,mm,dW,mR;
where:
e
– On (e=1), Off (e=0),
HH
– hour of the measurement start,
MM
– minutes of the measurement start,
hh
– hour of the measurement stop,
mm
– minutes of the measurement stop,
dW
– day of week in which the measurement will be done:
bit:0
– Monday,
...
bit:6
– Sunday
mR
– maximum number of the measurement days,
#7,AS, e,HH,MM,hh,mm,dW,mR;
where:
e
– On (e=1), Off (e=0),
HH
– hour of the measurement start,
MM
– minutes of the measurement start,
hh
– hour of the measurement stop,
mm
– minutes of the measurement stop,
dW
– day of week in which the measurement will be done:
bit:0
– Monday,
...
bit:6
– Sunday
mR
– maximum number of the measurement days,
Response format:
#7,AS;
Summary of Contents for SV 973
Page 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...