SVAN 973 User Manual
– Appendixes
143
y - depends of the detector type in selected profile: I (imp.), F (fast), S (slow) (cf. Tab. B.1.12)
3
y - only for exponential detector's type (cf. Tab. B.1.6)
(5) results of
1/1 OCTAVE
analysis or
1/3 OCTAVE
analysis if
1/1 OCTAVE
analysis or
1/3 OCTAVE
analysis was selected as the measurement function and the
LOGGER
was active
(paths:
Measurement
/
Logging
/
Logger Res.
/
Peak
Sp.
[√] and Leq Sp. [√]
);
the sequence of words is written:
<Octave Peak[1]> <Octave Peak [2]> ... <Octave Peak [Noct+NOctTot]> <Octave Leq[1]> <Octave
Leq[2]> ... <Octave Leq[NOct+NOctTot]>
where:
Octave Peak[i] - the result of
1/1 OCTAVE
or
1/3 OCTAVE
Peak analysis (*100 dB);
i = 1..NOct+NOctTot
Octave Leq[i] - the result of
1/1 OCTAVE
or
1/3 OCTAVE
Leq analysis (*100 dB);
i = 1..NOct+NOctTot
B.2.1.2. Record with the state of the markers
The record with the state of the markers consists of one word:
<0x8nnn>
in which 12 bits nnn denote the state of the markers:
b11 = state of #12 marker
b10 = state of #11 marker
...
b1 = state of #2 marker
b0 = state of #1 marker
B.2.1.3. Record with the breaks in the results registration
The record with the breaks in the results registration consists of four words:
<0xB0ii> <0xB1jj> <0xB2kk> <0xB3nn>
in which ii, jj, kk, nn bytes denote 4-bytes counter of left or skipped records: nnkkjjii (ii is the least significant
byte, nn
– the most significant byte).
B.2.1.4. Record with the breaks account PAUSE in the results registration
The record with the breaks in the results registration consists of four words:
<0xA0ii> <0xA1jj> <0xA2kk> <0xA3nn>
in which ii, jj, kk, nn bytes denote 4-bytes counter duration of PAUSE in milliseconds:
nnkkjjii (ii is the least significant byte, nn - the most significant byte).
B.2.1.5. Record with the wave file name
The record with the wave file name consists of six words:
<0xC2aa>
<0xccbb>
<0xeedd>
<0xggff>
<0xiihh>
Summary of Contents for SV 973
Page 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...