SVAN 973 User Manual
– Appendixes
165
Special filters
•
Windscreen
compensation filter improving the instrument frequency
response in the free acoustic field when windscreen SA22
is mounted on the microphone (see C.3)
RMS detector
•
Digital
“True RMS“ with Peak detection,
•
Resolution
0.1 dB
•
Range
327.7 dB
•
Crest Factor
unlimited (for signals in 8 kHz band).
Overload detector
The instrument has the built-in overload detectors. The overload in the measurement channel (in its analogue
part) and the overload of the analogue /
digital converter are both detected. The “overload” indication is when
the input signal amplitude
is 0.5 dB above
the declared “Peak measurement range”
Underrange detector
The instrument has the built-in under-
range detector. The “underrange” indication appears when the Leq
value for the elapsed time or the last second L
XY
value is below the lower linear operating range.
Time weighting characteristics
(Exponential averaging)
Slow
“
S
“ according to IEC 61672-1:2013 for Class 2, Equivalent Time Constant 1000 ms
Fast
“
F
” according to IEC 61672-1:2013 for Class 2, Equivalent Time Constant 125 ms
Impulse
“
I
” according to IEC 61672-1:2013 for Class 2, Equivalent Time Constant 35 ms, Hold Time 1500 s
Reference conditions
as per IEC 61252
•
Class of the acoustic field
Free field
•
Reference acoustic pressure
114.0 dB (related to 20
Pa)
•
Reference frequency
1000 Hz
•
Reference temperature
+20°C
•
Reference relative humidity
65 %
•
Reference static pressure
1013 hPa
•
Reference incidence direction
perpendicular to the microphone diaphragm.
Calibration
Acoustical
-
with
the SV 34B acoustic calibrator
(or equivalent):
•
Calibration level
114.0 dB (equal to the calibrator pressure level - see
calibration chart of the used calibrator)
Note:
The above levels correspond to 114 dB of calibrator’s sound pressure. If the calibrator
has a different sound pressure than 114 dB, the calibration levels must be accordingly adjusted.
Maximum peak voltage
3 V Peak-Peak (Maximum peak voltage of input sinusoidal
signal, which can be applied to the SLM without
destruction the meter)
Summary of Contents for SV 973
Page 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...