SVAN 973 User Manual
42
Speed of source value changing
Speed of triggering signal changing (
Gradient
) can be set in the range from
1 dB/ms
to
100 dB/ms
.
Recording before trigger
When the
Pre Trigger
parameter is switched on, the signal
will be recorded before the first trigger. The interval of such
recording is equal to the
Trigger period
(in the attached
example, 0.5 s).
►
Time of signal recording
The
Rec. Time
parameter
defines the time of signal recording after triggering. If
next trigger condition appears during the Recording Time, the signal will be
recorded for additional
Rec. Time
. The available values are from
1s
to
8h
, or
infinitive (
Inf
).
4.6 M
EASUREMENT RANGE
–
R
ANGE
The
Range
position
is used for checking the available
measurement range.
The absolute range values depend on the calibration
factor and are shown in the
Range
screen.
The attached screens assume the calibration factor equal
to zero.
<ENT>
The instrument uses two ranges
– one for the
Level Meter
,
1/1 Octave
,
1/3 Octave
and
RT60
functions (32.0 LEQ(A)
– 128.0 PEAK) and another for
the
Dosimeter
function (50.0 LEQ(A)
– 141.0 PEAK).
The detailed description of the measurement ranges
parameters is given in Appendix C.
Note: The calibration factor is always added to the lower and upper range level.
4.7 S
WITCHING ON COMPENSATION FILTERS
–
C
OMPENSATION
F
ILTER
The
Comp. Filter
position is available only in the
Advanced
interface mode and enables to switch on or off
compensation filters applied in the instrument.
The
Microphone
compensation filter (microphone inner
noise compensation) is switched on by default, however
you can switch it off for electrical measurements (e.g. for
laboratory calibration).
<ENT>
Summary of Contents for SV 973
Page 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...