131
SV 200A User Manual
EXTended I/O parameters
Word
number
Name
Comment
0
0xnn2E
[2E, nn=block’s length]
1
Mode
mode:
0 -
OFF
,
1 -
DIGITAL IN
,
2 -
DIGITAL OUT
2
Function
in the case of
DIGITAL IN
:
0 -
EXTERNAL TRIGGER
in the case of
DIGITAL OUT
:
0 -
TRIG. PULSE
,
1 -
ALARM PULSE
in other cases reserved
3
ActiveLevel
in the case of
DIGITAL OUT
and
ALARM PULSE
:
0 -
LOW
,
1 -
HIGH
in other cases reserved
4
Source
in the case of
DIGITAL OUT
and
ALARM PULSE
:
0 -
PEAK(1)
,
1 -
SPL(1)
,
2 -
LEQ(1)
in other cases reserved
5
AlarmLevel
in the case of
DIGITAL OUT
and
ALARM PULSE
:
30 ÷ 130 dB (*10)
in other cases reserved
6
Device
device connected to EXT I/O RS232 interface:
0 -
RS232 Interface
,
1 -
Meteo (SV205B)
2 -
reserved
3 -
Meteo (SV209 or SP275)
7
Reserved
reserved
8
Reserved
reserved
9
Polarisation/Slope
in the case of
DIGITAL OUT
and
TRIG. PULSE
:
Polarisation (0 -
POSITIVE
, 1 -
NEGATIVE
)
in the case of
DIGITAL IN
Slope (0 -
POSITIVE
, 1 -
NEGATIVE
)
in other cases reserved
...
Special settings for profiles
Word
number
Name
Comment
0
0xnn05
[05, nn=block’s length]
1
0x0307
[
used_profile, profile’s mask]
2
0xmm06
[06, mm=sub-
block’s length]
3
DetectorP[1]
detector type in the 1
st
profile:
0 -
IMP.
,
1 -
FAST
,
2 -
SLOW
4
FilterP[1]
filter type in the 1
st
profile:
1 -
Z
,
2 -
A
,
3 -
C