87
SV 200A User Manual
5.
Wait
until
instrument’s
settings
are
downloaded, which will be signalized by
Success
written in the last line of the
Reading
Settings…
box. Click OK to proceed.
6.
Station(s) configuration
window will appear. It allows the user to modify general instrument
settings, measurement setup, external I/O setup, internet services and system events settings.
Remote communication setup is available only for USB connection.
Note: The Station(s) configuration window enables configuring only general instrument
settings. The SVAN Files window with the use of Svan file editor function enables the
full range of settings
– see “+ User Manual”.
9.8.1 General settings
In the
General settings
tab, you can enter
Name
and
Comment
to the station. This can be useful while
using many measurement devices.
Note: Name appears on the Station list of the SvanNET web-service.
9.8.2 Measurement setup
The
Measurement setup
tab consists of the following parameters: measurement function (
Function
:
Level meter
,
1/1 Octave
or
1/3 Octave
), delay of the measurement start (
Start delay
), integration
period/measurement run time (
Integration period
:
INFINITE
,
1 s
÷ 24 h
), RMS detector type