85
SV 200A User Manual
3. view system log information (
Show log
button),
4. terminate the Live view session (
Terminate
button),
5. enable the SV
200A actuator and perform the instrument’s check (
Enable Actuator
button).
After enabling the actuator, the instrument starts measuring of the signal, generated by the actuator at
the level of 94 dB. If
Read step
is equal to
1s
then it is possible to observe the measured actuator level.
9.7 C
HANGING THE WORKING DIRECTORY
Working directory is a folder on the SD-card in which all the measurement files are stored.
Changing the working directory can be done with the
SVAN Files
feature of the
+
software
either via
SV 200 Instrument wizard
or via
Remote Communication Center
.
1. Make sure a measurement is not in progress.
2. Open the
SVAN Files
window and navigate to the desired directory using the left panel. The left
panel displays contents of the instrument
’s memory. Create a new directory if necessary.
SVAN Files button
SVAN Files button