59
SV 200A User Manual
Click
+Add alarm
in the STATIONS STATUS ALARMS pop-up box and new Alarm with CONDITIONS,
ACTIONS and MEASUREMENT POINTS areas will appear. Alarms are based on CONDITIONS and
relate to ACTIONS, that are default e-mails to the specified recipients, and refer to
MEASUREMENT POINTS. To configure Alarm:
1. click the
Status
button and in the EDIT CONDITIONS pop-up box:
a. select
Status source
:
Mains
,
External voltage
,
Battery charge
,
Storage memory
,
System check
or other positions,
b. click the
Trigger value
selector and choose the required threshold level for the selected
Status source
.
.
2. click
OK
and new condition will be displayed
in the CONDITIONS area.
3. click the
button to enter/edit e-mail recipients.
4. click the
Assign
button to refer alarm to the station(s).
1
2
a
b