SV 200A User Manual
4.4 C
ALIBRATION WITH THE USE OF
S
VAN
PC++
AND
USB
CONNECTION
1
Connect SV 200A
with the PC using SC 256A cable and start
+
.
2
When
SV 200A instrument wizard
appears on the screen, click
Calibration.
3
Set desired calibration level in the
SV 200A Calibration
window.
4
Attach the calibrator on the microphone, switch it on (if the used calibrator doesn’t have switch-on
automatic feature) and start the calibration measurement by pressing the
Start
button. The
successful calibration will result in calculation of the
Calibration Factor
. Calibration measurement
can be terminated by pressing the
Terminate
button.
Start button
Calibration level
Calibration button