75
SV 200A User Manual
Creating Actions
To create new action, click the
field and in
the
ADD EVENT ACTION
pop-up box, click the action
you wish to add and to configure:
Marker
,
Audio
,
I/O alarm
,
SMS alarm
or
E-mail alarm
.
After occurrence of the event, actions will be
performed during the time the event is active, at its
beginning or at the end depending on the action type.
The
Marker
action registers a marker in the logger file.
The Marker can be
Point
or
Block
type. The Point
marker means, that it will be registered only at the
beginning of the Event. The Block marker will be
registered at the beginning and at the end of the
Event.
The
Audio
action starts an audio recording in the
logger file (Event recording) or in the WAV type file
(Wave recording).
Max. duration
defines maximum time of recording
after the event start.
Min. break
defines the minimum time break between
two consequent records.
Pre trigger
enables earlier start of recording with
respect to the event start.
The period of such recording depends on the sample frequency and bits per sample. The maximum pre-
trigger period is:
•
for 24 bits per sample:
5s
for 48 kHz,
10s
for 24 kHz and
20s
for 12 kHz.
•
for 16 bits per sample:
8s
for 48 kHz,
15s
for 24 kHz and
30s
for 12 kHz.
The
I/O alarm
action starts an alarm signal at the
MULTI I/O socket output to which some alarm device
can be connected (for example, alarm lamp).
Max. duration
defines maximum time of alarm signal
after the event start.
Min. break
defines the minimum time break between
two consequent alarms.
The
SMS alarm
action sends the SMS note to the
defined
recipient’s phones, which are selected from
the
ADDRESS BOOK
.
Min. break
defines the minimum time break between
two consequent SMSs.
The SMS text is generated automatically in