110
SV 200A User Manual
Besides downloading files
AFD
performs other tasks such as:
•
checking station status and sending notifications when one of the instruments requires attention,
•
automatic time synchronization,
•
remote system check with in-built electrostatic actuator,
•
verification of instrument’s setups,
•
deleting already downloaded files,
•
publishing measurement data on the web server as a HTML file,
•
uploading measurement files on the FTP server.
Note: More information on Automatic Files Download can be found in + User
Manual: http://svantek.com/lang-en/support/19/svanpc_software.html#PDF.
Basic information
Automatic Files Download
runs as a Windows service. It does not require + application to
be running. Windows user who runs the
AFD
does not have to be logged on the PC.
AFD
automatically
resumes to its previous state after booting the PC.
The
AFD
window displays the list of stations running in the
AFD
mode. For selected instrument station
status information is available. It displays basic information (instrument’s serial number and download
path), downloads statistics and summary for last downloading, clearing, settings and system checks.
Information
section displays time left for downloading and time left for clearing. These values depend
on
Downloading period
and
Clearing period
settings, described below. Also, current state of the
AFD
module
is displayed.
Progress
section refers to the station selected in
station list
. It displays stations type and serial number
and the name of the file currently being downloaded.
Configuration
To configure
Automatic Files Download
session settings click
Options
button. Automatic files download options window will appear
allowing the user to make necessary adjustments.
Options
button is only active when the
AFD
is not running.
files to
download
download
period
automatic
cleaing
cleaing
period
system
validation
file saving settings
publishing
settings
measurement
checking