The
Marker
action means that
during alarm event the alarm marker
will be registered in the logger file at
the beginning and at the end of the
Event. The
Point marker
¸ which is
registered in the logger file only at
the beginning of the Event, can be
also selected.
The
Audio
action means that the
alarm event will trigger the audio
signal recording.
The audio signal recording can be
configured to start before the trigger
(
Pre trigger
) and stop after the
trigger
condition
disappears
(
Post trigger
).
Maximum duration
limits the audio
recording regardless of the event
duration.
Minimum
break
defines
the
minimum period of trigger inactivity
after start of the event. This
parameter damps so called pulse
alarms.
The
External I/O
action means that
the alarm event will switch on the
alarm signal generation on the
MULT. I/O
socket.
Maximum duration
limits the signal
generation period regardless of
event duration.
Minimum
break
defines
the
minimum period of signal trigger
inactivity after start of the event. This
parameter damps so called pulse
alarms.