99
SV 200A User Manual
Automatic system check
When the
Automatic system check
feature is enabled the calibration factor of the instrument will be
periodically verified using built-in electrostatic actuator. You can specify the hour and a weekday of the
automatic system checks.
To activate the automatic system check,
select
Enabled (no Cal. Factor update)
option from the
Mode
dropdown list.
To schedule automatic system checks,
choose weekday and select time of the system check from the
Do system check every day at (hh:mm)
dropdown list.
Note: System check cannot be considered as a calibration. Calibration factor will not be
updated during Automatic System Check procedure.
Note: Measurements are not stopped for the Automatic System Check procedure.
Calibration logging
Results of calibration (both manual and automatic) and automatic system check are stored in the
c
alibration and system check history file on the instrument’s SD-card. Log file name is
C.TXT
. The file
can be downloaded to the PC and then opened in any text editor, such as MS Notepad. Maximum
allowed size of log file is specified in the
Maximum calibration log file size [MB]
position.
9.8.12 Alarm settings
In the
Alarm configuration
section, press
Advanced Alarm Configuration
button to configure alarm
functionality of SV 200A. In the
Advanced Alarm Configuration
window, you can define and configure
the set of alarm
Conditions
and
Events
.
Alarm Conditions and
Events tabs
List of Alarm Conditions
Threshold conditions
configuration panel