100
SV 200A User Manual
9.8.12.1 Alarm Conditions
The
Conditions
tab enables configuring different types of alarm condition:
Threshold
,
Time
,
System
,
Spectrum
and
Meteo
. Already created alarm conditions are presented in the list on the right-side of the
Advanced Alarm Configuration
window. If you click on the specific alarm condition you will be able to
configure it in the right part of the window.
To create a new condition, click
Add
button and select a
new condition from the pop-down list.
The
Threshold
condition is based on comparison with the
Threshold
level such results as
Leq
,
Lpeak
,
Lmax
or
Lmin
(
Source
), calculated with weighting filter and LEQ detector for the selected
Profile
and
averaged during the
Integration
period. Two
Modes
are available:
Level+
and
Level-
. The Threshold
levels define the moment of the start/end of the condition. Additionally, you can set the
Minimum
Duration
time, which delays the condition start/end. The period when the condition is met is marked by
between-cursors of grey colour on the graph.
The
System
condition relates to system events, defined in the window below.
The
Time
condition is based on periods, defined for weekdays.
The
Spectrum
condition is based on comparison of spectrum of selected
Type
with the mask
–
threshold levels for octave bands.
System conditions
configuration panel
Time conditions
configuration panel