129
SV 200A User Manual
LOGGER TRIGGER parameters
Word
number
Name
Comment
0
0xnn2C
[2C, nn=block’s length]
1
TriggerMode
trigger mode:
0 -
OFF
,
4 - measurement on trigger
LEVEL+
,
5 - measurement on trigger
LEVEL
–
2
TriggerSource
source of the triggering signal:
0 -
RMS(1)
the
RMS
result from the first profile
3
TriggerLev
level of triggering:
25 ÷ 130 dB (*10)
4
TriggerGrad
reserved
5
TriggerPre
number of the records taken into account before the fulfilment of
the triggering condition
(1
50)
6
TriggerPost
number of the records taken into account after the fulfilment of
the triggering condition
(1
200)
7
TriggerSampling
reserved
8
TriggerRecTime
reserved
9
TriggerStep
reserved
10
TriggerFilter
reserved
11
TriggerBitsPerSample
reserved
12
Range
reserved
…
Time-domain signal recording parameters
Word
number
Name
Comment
0
0xnn31
[31, nn=block’s length]
1
TriggerMode
trigger mode:
0 -
OFF
,
1 - recording whole measurement
2 - recording on trigger
SLOPE+
3 - recording on trigger
SLOPE
–
4 - recording on trigger
LEVEL+
5 - recording on trigger
LEVEL
–
6 - recording on trigger
GRAD+
8 - recording on
Integration Period
trigger
9 - recording on
Remote Command
trigger
2
TriggerSource
source of the triggering signal:
0 -
RMS(1)
the
RMS
result from the first profile
in the case of TriggerMode=
SLOPE+
or
SLOPE
–
:
1 -
External IO
3
TriggerLevel
level of triggering:
25 ÷ 130 dB (*10)
4
TriggerGrad
gradient of triggering:
1 dB/ms ÷ 100 dB/ms (*10)
5
TriggerPre
pre-trigger time given in 10ms
6
TriggerPost
reserved
7
TriggerSampling
sampling frequency in 10 Hz