Instruction manual for Evario hospital bed
150
15.5 Fault Indication on the Control Box and Locking Box
If your bed is fitted with this equipment, it is easy to obtain a quick overview of the functional status
and to perform a RESET (to acknowledge error messages) if necessary.
For safety reasons, if a serious error is detected, all functions are locked electronically.
Any faults that are recognised are displayed by the locking LEDs flashing on the control
box/locking box, making it easier to rectify the fault
The table below shows the fault indications on the control box and the locking box as well as pos-
sible solutions.
A2 B1 B2 C1 C2 D1
D2
Acoustic
signal
when
button
pressed
Error
Solution
X
X
Thigh rest motor (socket 2) limit
switch
Check / replace
cable + plug-in
connections
+ acknowledge
fault (RESET;
Chapter 15.6)
More rarely: Re-
place the motor
X
X
Mattress base height motor limit
switch, head end (socket 1)
X
X
X
Position sensor for mattress
base height motor, head end
(socket 1)
X
X
Mattress base height motor limit
switch, foot end (socket 3)
X
X
X
Position sensor for mattress
base height motor, foot end
(socket 3)
X
X
X
X
Fault in a control unit (e.g.
jammed button, broken cable),
more rarely: internal fault in the
control unit
Check each con-
trol unit, replace
any faulty control
unit or cable
+ acknowledge
fault (RESET;
Chapter 15.6)
More rarely: Re-
place control unit
Fig. 32: Fault indications on locking box
Summary of Contents for Evario
Page 26: ...Instruction manual for Evario hospital bed 26 Fig 3 Positions for storing the LCD handset...
Page 125: ...Instruction manual for Evario hospital bed 125 22 22...
Page 166: ...Instruction manual for Evario hospital bed 166 Notes...
Page 167: ...Instruction manual for Evario hospital bed 167...
Page 168: ...Australian Distributor...