Instruction manual for Evario hospital bed
140
14.10 P
rogramming the LCD Handset
It is necessary to programme the LCD handset in the following cases:
x
If you are connecting a different handset to the existing control unit,
x
If a new control unit is connected.
During programming, the LCD handset recognises the range of functions of the control unit
and is adapted accordingly. The LCD handset will not work unless it has been programmed!
IMPORTANT
The LCD handset must only be programmed by technical staff or qualified medical staff who have
been trained by the operator to do so.
1 The warning triangle symbol is displayed
on the LCD handset. All adjustment op-
tions are locked.
2 Press the UP and DOWN buttons simulta-
neously and at the same time hold the top
end of the LCD handset briefly against the
magnetic unlocking key.
3 Keep both the UP and DOWN buttons
pressed for a further 5 seconds until OK !
is displayed.
4 Select the RESET function by pressing the
toggle button.