Instruction manual for Evario hospital bed
134
14.7 Replace the Control Box / Locking Box
1
Unplug the electricity cable from the electrical socket.
2
Remove the retaining ring on the connection socket from the control box / locking box and
pull out the OpenBus® connector
3
Replace the control box / locking box. Insert the OpenBus® plug back into the open socket
as far as it will go. Make sure that the plug is correctly aligned (plug groove aligned with
the protrusion in the socket) and that the O-ring on the plug is not damaged. This ring en-
sures that the plug is tightly sealed.
4
Re-attach the retaining ring. Programme the control box / locking box (RESET - see Chap-
ter 15.6
»
5
Test the power adjustments to ensure they are working correctly!