15.2 Fault Display on LCD Handset/Outside Control Panel
Faults in the adjustment drives and the operating units are indicated on the display.
Indication on LCD
handset
Type of fault
Solution
General fault indication on the dis-
play. Adjustment drive cannot be
activated. An acoustic signal
sounds when a button is pressed.
1.Start the first step A or B of the
RESET procedure to display the
exact error image according to the
troubleshooting guide (see Chap-
te2.Resolve the fault according to
the troubleshooting guide (see
Chapter 15.3 ).
3. Conduct the entire RESET pro-
cess completely to acknowledge
the fault (see Chapter 15.3)
Malfunction of a handset / control
unit (e.g. jammed button, broken
cable)
Id replace defective LCD
handset / control unit + RESET
Internal control unit fault (very rare)
Replace control unit
Height adjustment fault (actuator at
head end)
1. Check/replace ca-
bles/plugs + RESET + ini-
tialise
2. If error still persists: Re-
place motor + RESET +
initialise
Height adjustment fault (actuator at
head end)
Height adjustment fault in both
actuators
Backrest actuator fault
Thigh rest actuator fault