15.3 RESET: Acknowledging Faults on LCD Handset / Control Panel
For safety reasons, if a serious error is detected by the control unit all functions are electronically
locked for safety reasons.
WARNING
Risk of injury
If the control unit automatically locks again within a short space of time, contact
Stiegelmeyer’s Customer Service to rectify the cause of the fault.
Ignoring this safety cut-out function, for instance, by repeatedly resetting the control unit
without rectifying the cause of the fault, may result in endangering the patient due to in-
advertent actuation of the control unit.
Depending on the bed fixtures, the locked function is indicated by:
an appropriate warning symbol
the display of connected control units
the locking buttons concerned on the locking box/control box will alternately flash green and
orange.
If the patient presses a button, a signal tone indicates a fault.
If after a RESET, it is still not possible to carry out a mattress base height adjustment, please addi-
tionally carry out a new initialisation.
IMPORTANT
Faults must only be rectified by service personnel.
Rectify faults as instructed in the fault rectification table. Faults must not be
acknowledged until this has been done.
Alternatively, RESET can also be carried out very easily via the control box / locking box
(optional) instead of on the LCD handset/ control panel (see Chapter 15.6).
A: Start on LCD Handset:
Press the UP and DOWN buttons simultaneously
und and at the same time hold the top end of the LCD handset against
the orange magnetic unlocking key
,
until the warning symbol
disappears and the current error im-
age appears (see Troubleshooting Guide)
or
B: Start on Control Panel:
Press the
button until the warning symbol
disappears and
the current error image appears (example: Symbol for height adjust-
ment drive fault; see Troubleshooting Guide)
The procedure after this is identical in principle for both control units A + B:
!
Summary of Contents for Evario
Page 26: ...Instruction manual for Evario hospital bed 26 Fig 3 Positions for storing the LCD handset...
Page 125: ...Instruction manual for Evario hospital bed 125 22 22...
Page 166: ...Instruction manual for Evario hospital bed 166 Notes...
Page 167: ...Instruction manual for Evario hospital bed 167...
Page 168: ...Australian Distributor...