Instruction manual for Evario hospital bed
148
3 Select the initialisation function by press-
ing the toggle switch.
4 Press the UP button.
5 The last error function will appear on the
display.
6 Press the UP button again to display the last
fault. The symbol for the last fault that oc-
curred will be displayed (Example: Symbol
for height adjustment drive fault).
Press any button to return to the function selector menu.
7 Select the CLEAR function by pressing
the toggle switch.
8 Press the UP button to confirm that the fault
should be deleted.
9 OK ! confirms the deletion.