12.3.1.2 On the Control Panel (Outside)
1 Press the settings button. The CPR sym-
bol appears on the display.
2 Select the washing position adjustment
function using the toggle button.
3 Press the DOWN or UP button. The bed
moves to the washing position.
12.3.1.3 On the Control Box
12.3.2 Defining an Alternative Washing Position
The height and angle of tilt of the bed in the factory default washing position can be adapted to the
particular washing system used. The backrest and thigh rest always move into the lowest position.
IMPORTANT
To ensure that cleaning fluids can drain off unhindered, the bed must not be washed in a horizon-
tal or reverse-Trendelenburg position.
The bed settings to be established as the washing position in the washing machine are:
Trendelenburg position (angle of tilt at least 10°),
All rests set to the lowest position,
All safety sides manually raised.
1 Press the unlock button and the Trendelenburg position button
until the full extent or the desired position is reached.