Instruction manual for Evario hospital bed
127
13.6 Servicing Plan
Components
Action
Interval
Information
All screw connections
and/or metal safety
caps
Check that they are
firmly positioned; tighten
and/or replace if neces-
sary.
At least once a year
Use suitable tools.
Connecting bolts on
the removable head-
board and footboard
Clean and slightly
grease, if necessary.
At least once a year
Use resin-free and
acid-free grease (e.g.
grease C).
If machine washed,
at least after every
25 washes. Clean-
ing
Pivot and friction bear-
ings
Clean and lightly spray.
As necessary
Use resin-free and
acid-free oil spray
(e.g. lubricant A).
Bowden cable for CPR
lowering of backrest
Check that they are
correctly adjusted and
routed.
At least once a year
There must be no
sharp bends or kinks!
Lead-acid batteries
Check whether charged
and ready for use.
We recommend that
lead-acid batteries
are replaced after 5
years at the latest
(wear and tear part)
Connect bed to
mains supply.
Li-ion batteries
Check whether charged
and ready for use.
We recommend that
rechargeable batter-
ies are replaced
after 5 years (wear
and tear part).
Connect bed to
mains supply.
Earth wires
Check for tears and
tightness of screws.
At least once a year
Replace torn or dam-
aged earth wires.
Damaged coating
Touch up any damaged
areas of the coating:
As necessary
Special-purpose
paint (mechanically
stable anti-corrosive
protection) in a suita-
ble colour.
O-rings ("sealing
rings") on the electric
plugs
Always check whether
they are present and/or
damaged.
Whenever plugs
have been discon-
nected (e.g. after
replacing actuators
or during trouble-
shooting)
The plugs of new
components are al-
ways fitted with a
new O-ring.
Replace O-rings with original LINAK O-rings only. Check the corresponding recesses in
the plug for dirt or damage, and lubricate with Vaseline, before reinserting the plugs.
We recommend supplementing the specified maintenance work with appropriate preventive
measures to further safeguard trouble-free and fail-safe operation of the bed. A guide to preventive
replacement of potentially relevant components can be obtained from our service centre:
Stiegelmeyer GmbH & Co. KG
Tel.: +49 (0) 5221 185-
777 •
Fax: +49(0) 5221 185-219
Email:
Summary of Contents for Evario
Page 26: ...Instruction manual for Evario hospital bed 26 Fig 3 Positions for storing the LCD handset...
Page 125: ...Instruction manual for Evario hospital bed 125 22 22...
Page 166: ...Instruction manual for Evario hospital bed 166 Notes...
Page 167: ...Instruction manual for Evario hospital bed 167...
Page 168: ...Australian Distributor...