Error
Code
Error Message
Description
48
Emergency Stop
Pressed
Determine reason why emergency stop was pressed. Reset
emergency stop by pulling the button to release, and
confirming on the display.
49
Micro-controller error
Switch the equipment off, then on again. If problem persists,
contact supplier to arrange a qualified technician inspection.
50
Internal ambient
temperature too high
Make sure room temperature is below +40 °C (+104 °F).
Make sure cooling fan air intake is not blocked or restricted.
Turn equipment off and leave for 1 hour to cool down.
If problem persists, contact supplier to arrange a qualified
technician inspection.
51
Fill level sensor
invalid
Repeat cycle.
If problem persists, contact supplier to arrange a qualified
technician inspection.
52
Real-time clock read
error
Switch the equipment off, then on again.
If problem persists, contact supplier to arrange a qualified
technician inspection.
53
24V power supply
voltage out-of-range
Switch the equipment off, then on again.
If problem persists, contact supplier to arrange a qualified
technician inspection.
54
Overfill occurred
during cool down
process
Make sure that the filter is clean and not clogged. See
step
2
of
Cleaning and Inspection Before Every Wash Cycle
Make sure that the drain hose is not kinked or constricted,
and that site drain is not blocked.
Repeat cycle.
If problem persists, contact supplier to arrange a qualified
technician inspection.
Table 12.1
Error Codes and Remedy Actions (continued)
TROUBLESHOOTING
147
InnoWave Pro Sonic Irrigator
Operator Manual
Summary of Contents for InnoWave Pro
Page 1: ...InnoWave Pro Sonic Irrigator Operator Manual ENGLISH ...
Page 2: ......
Page 10: ...viii This page intentionally blank InnoWave Pro Sonic Irrigator Operator Manual ...
Page 22: ...12 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 30: ...20 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 96: ...86 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 112: ...102 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 118: ...108 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 134: ...124 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 146: ...136 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 150: ...140 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 170: ...160 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 176: ...166 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 192: ...182 APPENDICES Operator Manual InnoWave Pro Sonic Irrigator This page intentionally blank ...
Page 194: ...184 APPENDICES Operator Manual InnoWave Pro Sonic Irrigator This page intentionally blank ...
Page 196: ...186 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...