Error
Code
Error Message
Description
14
Chamber start
temperature too high
Tank has not cooled sufficiently since last wash. Make sure
cold water supply is below +35 °C (+95 °F).
Make sure that the cold water inlet hose is connected and
not kinked or blocked.
Make sure that the site cold water supply is on.
Allow more time between washes for the equipment to cool
down.
If problem persists, contact supplier to arrange a qualified
technician inspection
15
Tank temperature
out of range
Switch the equipment off then on again. If problem persists,
contact supplier to arrange a qualified technician inspection.
16
Bottom level sensor
invalid
Switch the equipment off then on again.
If problem persists, contact supplier to arrange a qualified
technician inspection.
17
Bottom or minimum
level sensor invalid
Switch the equipment off then on again.
If problem persists, contact supplier to arrange a qualified
technician inspection.
18
Minimum level
sensor invalid
Switch the equipment off then on again.
If problem persists, contact supplier to arrange a qualified
technician inspection.
19
Overflow level sensor
invalid
Switch the equipment off then on again.
If problem persists, contact supplier to arrange a qualified
technician inspection.
20
User aborted cycle
Determine reason why user aborted cycle.
Repeat cycle.
Table 12.1
Error Codes and Remedy Actions (continued)
144
TROUBLESHOOTING
InnoWave Pro Sonic Irrigator
Operator Manual
Summary of Contents for InnoWave Pro
Page 1: ...InnoWave Pro Sonic Irrigator Operator Manual ENGLISH ...
Page 2: ......
Page 10: ...viii This page intentionally blank InnoWave Pro Sonic Irrigator Operator Manual ...
Page 22: ...12 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 30: ...20 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 96: ...86 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 112: ...102 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 118: ...108 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 134: ...124 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 146: ...136 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 150: ...140 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 170: ...160 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 176: ...166 InnoWave Pro Sonic Irrigator Operator Manual This page intentionally blank ...
Page 192: ...182 APPENDICES Operator Manual InnoWave Pro Sonic Irrigator This page intentionally blank ...
Page 194: ...184 APPENDICES Operator Manual InnoWave Pro Sonic Irrigator This page intentionally blank ...