5-7
Routine Maintenance
Operator Manual
150830-685
5.6 Preventive
Maintenance Schedule
Maintenance procedures described in
through
5.5
must
be performed regularly at the indicated intervals. Use the Preventive
Maintenance Schedule in
as a guide for further Cmax
Surgical Table maintenance. Table usage may require more frequent
maintenance than indicated. See
R
EPLACEMENT
P
ARTS
.
Customer should maintain a record of all maintenance procedures
performed on the table.
If a problem occurs, refer to
NOTE: Never permit unqualified persons to service the table and
Preventive Maintenance is not covered under warranty.
WARNING – PERSONAL
INJURY AND/OR EQUIPMENT
DAMAGE HAZARD: Failure to
perform periodic inspections
of the table could result in
serious personal injury or
equipment damage.
Table 5-2. Preventive Maintenance Schedule for Cmax Surgical Table
Service Required
Minimum Frequency
1.0 PREPARATION FOR PREVENTIVE MAINTENANCE
Discuss equipment operation with department personnel.
1 x per year
Remove pads. Examine pad covers and fastening strips (pads and table).
1 x per year
Examine clamps and other side rail hardware
.
1 x per year
Tighten side rails.
1 x per year
Unfasten and remove table shrouds and base cover.
1 x per year
Remove and inspect x-ray tops on each tabletop section (if equipped).
1 x per year
2.0 HYDRAULIC SYSTEM
Check hydraulic oil level.
1 x per year
Inspect floor directly beneath table and all hoses, fittings and components of
hydraulic system for evidence of oil leaks.
1 x per year
Verify operation of Auxiliary Control System foot pedal.
1 x per year
3.0 CASTERS AND FLOOR LOCKS
Clean and inspect casters.
1 x per year
Lubricate casters.
1 x per year
Check floor-lock mechanism for proper operation.
1 x per year
4.0 CONTROLS
Using hand control or optional foot control, check table positioning.
1 x per year
5.0 ELECTRICAL CHECKS
Ensure all circuit board connectors and cable plugs are tight.
1 x per year
Check all cables for damage or fraying.
1 x per year
Inspect electrical cord and hand control cable.
1 x per year
Verify operation of Auxiliary Control System.
1 x per year
Summary of Contents for Cmax
Page 1: ......
Page 4: ......
Page 65: ......
Page 142: ...Page 3 of 14 ...
Page 148: ...Page 9 of 14 ...
Page 149: ...Page 10 of 14 ...
Page 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Page 164: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......