Table 2. Universal Hand Control and Surgical Table Compatibility Modes Do Not Match
Message Display With
Numeric Error Codes
Universal Hand Control displays numeric error messages beside
Wrench
symbol when V2 table control board is installed in a Surgical Table.
Numeric error codes persist during a key press for articulation related
errors. Numeric error codes persist until the Surgical Table is powered off
for errors related to system failures. Figure 8 shows a typical error code
display.
Table 3
lists error codes (wrench codes) and their meanings.
Figure 8. Typical Error Code Display - Error Code 20 Displayed
Compatibility
Mode
Table Type
Universal Hand Control And Surgical
Table Behavior
Corrective Action
Cmax/4085
5085
When the 5085 is not in Beach Chair Mode,
Universal Hand Control and 5085 Surgical
Table function correctly.
Refer to Operator’s
Manual
symbol appears in lower right corner
of display. When table is powered on again,
user is prompted to select 5085 Compatibility
Mode. When Beach Chair Mode is engaged,
Universal Hand Control displays
BEACH
CHAIR MODE ERROR
(see BEACH CHAIR
MODE for 5085 Section and Figure 10)
.
Power table off by
pressing
STOP
button.
Power table on by
pressing any key
except
STOP
and
select 5085 Mode
when prompted.
5085
Cmax/4085 With
Version 2 Table
Control Chassis
Universal Hand Control and Surgical Table
function correctly.
Refer to Operator’s
Manual
symbol appears in lower right corner
of display. When table is powered on again,
user is prompted to select 4085 Compatibility
Mode.
Power table off by
pressing
STOP
button.
Power table on by
pressing any key
except
STOP
and
select 4085 Mode
when prompted.
5085
Cmax/4085 With
Version 1 Table
Control Chassis
Running Version
1.46 or Higher
Software
Universal Hand Control display shows
BAD
VERSION
message but otherwise Hand
Control and Surgical Table function correctly.
Refer to Operator’s Manual
symbol appears
in lower right corner of display. When table is
powered on again, user is prompted to select
4085 Compatibility Mode.
Power table off by
pressing
STOP
button.
Power table on by
pressing any key
except
STOP
and
select 4085 Mode
when prompted.
5085
Cmax/4085 With
Version 1 Table
Control Chassis
Running Version
1.43 Software
Universal Hand Control and Surgical Table
shut down after approximately seven seconds.
Power table on by
pressing any key
except
STOP
and
select 4085 Mode
immediately when
prompted.
Summary of Contents for Cmax
Page 1: ......
Page 4: ......
Page 65: ......
Page 142: ...Page 3 of 14 ...
Page 148: ...Page 9 of 14 ...
Page 149: ...Page 10 of 14 ...
Page 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Page 164: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......