the button and the table control software version will be displayed.
1) If the table control version is currently
v1.48
, then the table
control board is a 542200-903 and will need to be replaced with
P150832-769. Once replaced proceed to number 2.
2) If the table control version is currently
v2.04
or
v2.05
, then the table
control board will need to be updated to software version
2.07
. This
also indicates you have Master Control Board P150832-769. Follow
Cmax Service Diagnostics and Application Download Instructions
P/N P764333-920, also found on the Equipment Software Downloads
Page to update the software.
Trend/Tilt Sensor Replacement
1.
Remove existing sensors
.
a. Both the Trend and the Tilt sensors will need to be removed. The new
sensor is a dual axis sensor and functions as both the Trend/Tilt sensor.
b. Follow the Maintenance Manual Instruction (Cmax – P764333-268 or 4085-
P764333-765) for Trend/Tilt Sensor Replacement to gain access to the
sensors. The existing bracket containing the Trend Sensor is located on the
patient left side of the table.