5-3
Routine Maintenance
Operator Manual
150830-685
5.2.2 After Each Usage
After each use of the Cmax Surgical Table, clean/disinfect as
follows:
NOTE: Ensure all protective covers are installed over any open
receptacles.
1.
Remove gross soil with a disposable cloth and place used cloth
in an appropriate biohazardous waste disposal container.
2.
Clean tabletop and pads as follows:
a.
Articulate tabletop to level and place at comfortable working
height.
b. Gently "peel" pad away from tabletop fastening strips and
place on another table or flat surface.
c.
Hold can 6-8" (152-203 mm) from surface and spray
cleaning fluid liberally on top and sides of pads. (It is
recommended to clean only one pad at a time.) Carefully
avoid any pad seams.
NOTE: Do not use any oxidizing products with chloric deriva-
tive base such as bleach, or oxygenic derivatives such as
peracetic acid.
d. Wipe sprayed surfaces with a clean lint-free cloth dampened
with water to remove cleaning fluid (dampening cloth will
minimize streaking).
e.
Wipe cleaned surfaces again with clean, damp, lint-free
cloth to remove residue.
f.
Wipe cleaned surfaces again with clean, dry, lint-free cloth
to remove all moisture.
g. Repeat
Steps c, d, and e
for other pads. Repeat
Steps c, d,
e, and f
for tabletop surfaces exposed when pads are
removed. Thoroughly dry bottom of pads and tabletop
surfaces.
h.
Place pads back onto tabletop by aligning with sides of table
and pressing into place on fastening strips.
3.
Raise table to maximum height to access lower surfaces.
4.
Clean yoke, shrouds and entire base surface as follows:
a.
Hold can 6-8" (152-203 mm) from surface and spray
cleaning fluid liberally on yoke and shrouds. Avoid spraying
in direction of connectors.
b. Wipe sprayed surfaces with a clean cloth dampened with
water to remove cleaning fluid (dampening cloth will
minimize streaking).
c.
Wipe cleaned surfaces again with a clean, damp, lint-free
cloth to remove residue.
d. Repeat
Steps a, b, and c
for base surface.
WARNING – INFECTION
HAZARD: To protect against
aerosols being reflected from
potentially contaminated
surfaces, wear rubber or
plastic gloves, masks and eye
protection, and follow OSHA
blood-borne pathogens
standards when cleaning.
WARNING –PERSONAL
INJURY HAZARD: When
cleaning/disinfecting table, do
not use phenolics, which may
cause patient skin burns if
inadequately rinsed off, or
alcohol, which does not have
sufficient cleaning/
disinfection properties.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE:
• When cleaning/disinfecting
table, thoroughly read the
cleaning fluid directions for
use and follow all
directions and cautions as
shown.
• Cleaning procedures
requiring articulation of the
table should be performed
only by persons familiar
with table operation.
• Do not spray cleaning fluid
into electric receptacles
and avoid spraying directly
on emergency backup
buttons or into clearance
space. Spray or drippage
may settle onto electric
circuits inside table
causing corrosion and loss
of function.
• After performing cleaning
procedures, ensure pads,
tabletop and x-ray tops are
completely dry before
reinstalling. Moisture
trapped between pads and
x-ray tops may contribute
to equipment damage, such
as x-ray top warpage.
Summary of Contents for Cmax
Page 1: ......
Page 4: ......
Page 65: ......
Page 142: ...Page 3 of 14 ...
Page 148: ...Page 9 of 14 ...
Page 149: ...Page 10 of 14 ...
Page 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Page 164: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......