2-3
Important User Information
Operator Manual
150830-685
2.2 Patient Positioning
and Weight Limitation
NOTE: Accessories may have a specified weight limitation that is less
than that of the table. Do not exceed the lowest weight limit, table or
accessory.
The Cmax General Surgical Table is designed to safely support (with
patient in normal orientation): 1100 lb (499 kg) without patient
posturing or tabletop slide (except raise/lower); 1000 lb (454 kg) full
patient posturing (except tabletop slide); 600 lb (272 kg) full patient
posturing (including tabletop slide).
Do not exceed the maximum
patient weight.
NOTE: When positioning the patient on the table, note the following:
1) Always check patient stability when patient is positioned.
2) Do not place patient on the table unless floor locks are
engaged.
3) Use extreme care when transferring patients to or from table.
4) Ensure all accessories are properly installed and secured.
5) Check for and eliminate harmful patient pressure points once
patient is positioned.
6) Have a qualified medical professional monitor patient during
surgery for all possible patient positioning hazards.
7) Do not place patient weight on the Leg Section. Leg Section is
designed to support leg weight, not body weight.
All patients must be restrained for proper safety regardless of the
length or type of procedure.
NOTE: STERIS recommends only using STERIS manufactured or
distributed accessories with this table. If an accessory from another
manufacturer is used, it is the user’s responsibility to ensure the
accessory is compatible with the table and not dangerous to the patient
and/or surgical staff.
WARNING – TIPPING
HAZARD:
• Do not place patient on the
table unless floor locks are
engaged.
• Do not use this table for
patients exceeding the
maximum patient weight
(patient in normal
orientation) of 1100 lb
(499 kg) without patient
posturing or tabletop slide
(except raise/lower), 1000 lb
(454 kg) with patient
posturing (except tabletop
slide), or 600 lb (272 kg)
with full patient posturing
(including tabletop slide).
WARNING – PERSONAL
INJURY HAZARD:
• The floor locks must be
engaged at all times when a
patient is on the table.
Failure to engage the floor
locks before placing a
patient on the table or
disengaging the floor locks
while a patient is on the
table could result in the
table rolling unexpectedly
during the procedure.
• Healthcare professionals
must ensure patients are
positioned and monitored
so as to prevent compro-
mising respiration, nerve
pathways or circulation
.
Summary of Contents for Cmax
Page 1: ......
Page 4: ......
Page 65: ......
Page 142: ...Page 3 of 14 ...
Page 148: ...Page 9 of 14 ...
Page 149: ...Page 10 of 14 ...
Page 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Page 164: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......