4-9
Operating Instructions
Operator Manual
150830-685
4.3 Reverse Patient
Orientation
Some surgeries (e.g., shoulder) require the patient to be in reverse
positioning and the table to be operated in a reverse configuration.
Press
REV. ORIENT
button on hand control (see Figure 4-3), LED
should light. Press
REV. ORIENT
button again to cancel Reverse
Orientation function.
When
REV. ORIENT
button is pressed, all other hand control
commands (through the control system) note the new patient head
position. Trendelenburg and Lateral Tilt movements and Back and
Leg Sections are automatically reversed.
IMPORTANT:
When using the
REV. ORIENT
button, note the
following:
1) The information displayed on the LCD is NOT reversed.
2) When table is turned OFF, control system stores orientation
status. Before any new positioning is attempted, ensure table
patient configuration is correct.
3) The Cmax Table Head Section is manually adjusted and will
not be affected by
REV. ORIENT
button operation.
4) FLEX and REFLEX functions are NOT available for patient
safety.
5) When operating in REVERSE ORIENTATION,
LEVEL
function
does NOT operate.
6) The hand control automatically deactivates after two minutes of
nonuse. The hand control is reactivated by depressing
ANY
button but
OFF.
4.4 Optional Foot
Control Operation
The Cmax Surgical Table may be operated using the optional foot
control as follows:
1.
Connect foot control cord to appropriate control panel socket on
table column (refer to Figure 3-5).
2.
The foot control allows operator to control HEIGHT, TREND, and
TILT movements.
3.
Press and release any positioning pedal to switch table control
ON. If the hand control is also attached, PEDAL ON will appear
on the LCD Display message window.
NOTE: When using the hand control and optional foot control at the
same time, the hand control has priority. Also, if the hand control
OFF
button is pressed, foot control will also switch OFF.
4.
If foot control is unplugged from table, ENSURE protective cover
is placed over opening.
4.5 Auxiliary Control
System
The Cmax Surgical Table is equipped with an auxiliary control
system. This system can be actuated at any time and will allow table
operation in the event of primary control malfunction. A pedal (foot
pump to provide hydraulic power) and override hand control (see
Figure 4-4 and Figure 4-6) are located on the right side of the table
base behind a cover panel and are used for table movements.
NOTE: The Cmax Surgical Table head section must be manually
adjusted.
Summary of Contents for Cmax
Page 1: ......
Page 4: ......
Page 65: ......
Page 142: ...Page 3 of 14 ...
Page 148: ...Page 9 of 14 ...
Page 149: ...Page 10 of 14 ...
Page 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Page 164: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......