6-2
150830-685
Operator Manual
Troubleshooting
* Service charges may be incurred. Consult your warranty for details.
** Use optional foot control or auxiliary (override) control system if hand control is faulty [refer to
].
Table 6-1. Troubleshooting Chart
SITUATION
ACTION
Hand Control does not activate when table is plugged
into electrical outlet and any pushbutton (except
OFF
)
is pressed .
Main battery fuse blown – replace.
Faulty power cord – replace.
Hand control not plugged in properly – correct.
Faulty hand control** – replace.
Call STERIS.*
Table remains on battery power even when main
power cord is plugged into facility power.
Incorrect power cord connection – correct.
Faulty power cord – replace.
Main fuse blown – replace.
Call STERIS.*
No functions using hand control.
Hand Control not plugged in properly – correct.
No Power – table deactivates after two minutes of
non-use. Press any button on hand control (except
OFF
).
Faulty hand control** – replace.
Call STERIS.*
Table will not tilt right/left using hand control.
Hand control not on – Press any button on hand
control (except
OFF
).
Faulty Hand control** – replace.
Call STERIS.*
Table will not Trendeleburg/Reverse Trendeleburg
using hand control.
Hand control not on – Press any button on hand
control (except
OFF
).
Faulty Hand Control** – replace.
Call STERIS.*
Table will not raise/lower using hand control.
Hand control not on – Press any button on hand
control (except
OFF
).
Faulty hand control** – replace.
Call STERIS.*
Floor locks will not release and SLIDE HEAD or SLIDE
FOOT message is displayed.
Move SLIDE to 0”.
Summary of Contents for Cmax
Page 1: ......
Page 4: ......
Page 65: ......
Page 142: ...Page 3 of 14 ...
Page 148: ...Page 9 of 14 ...
Page 149: ...Page 10 of 14 ...
Page 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Page 164: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......