2-4
150830-685
Operator Manual
Important User Information
2.3 Patient Safety
Straps
Adjust the patient safety straps such that they restrain the patient
according to currently accepted patient restraining practices, keeping
in mind the Trendelenburg/Reverse Trendelenburg, back, lateral tilt
and height adjustment.
2.4 General
Description
The Cmax Surgical Table (see Figure 2-3) is a mobile,
electrohydraulically operated surgical table specifically designed to
provide the complete patient positioning flexibility required for modern
surgical care facilities. The Cmax Surgical Table features powered
lateral tilt, Trendelenburg/Reverse Trendelenburg and adjustable
height functions. This table is designed to safely function with patients
not exceeding the maximum patient weight limit (patient in normal
orientation) of 1100 lb (499 kg) without patient posturing or tabletop
slide (except raise/lower); 1000 lb (454 kg) full patient posturing
(except tabletop slide);600 lb (272 kg) full patient posturing (including
tabletop slide) and is constructed of aluminum alloy, stainless steel
and other high quality materials. The Cmax Surgical Table is
equipped with a large sliding table top providing unrestricted
radiological access without patient reversing.
The Cmax Surgical Table is powered by either internal battery or
facility electric through use of the patent-pending INTELLIPOWER
®
dual power system.
The table accepts positioning commands from three sources:
1.
A
hand control
(which includes “FLEX”, "REFLEX", "SLIDE HEAD",
"SLIDE FOOT", "OFF", "Trend.”, "REV. TREND", “Height UP”,
"HEIGHT DOWN", “TILT LEFT”, "TILT RIGHT", "BACK DOWN",
"BACK UP", "LEG DOWN" , "LEG UP", "KIDNEY UP", "KIDNEY
DOWN", "LEVEL", "REV. ORIENT" and "FLOOR LOCK". Refer to
Figure 4-3 of this document for control feature locations.
2.
An optional physician-controlled
foot control
(which includes
Trendelenburg Tilts, Lateral Tilts and Height functions). Refer to
.
3.
An
Auxiliary Control System
permits table operation in the event
of primary control malfunction (refer to
).
NOTE: The Head Section is manually adjustable.
* SAE Measurements are approximate based on metric dimensions.
WARNING – PERSONAL
INJURY HAZARD: Failure to
keep the patient properly
secured with the patient
safety straps at all times could
result in death or serious
injury to the patient and the
operating room staff
.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE:
Appropriate components of
this surgical table have been
tested and found in
compliance with IEC 60601-1-
2:2001, Medical Electrical
Equipment - Part 1; General
Requirements for Safety;
Electromagnetic Compatibility
(EMC). There is, however, a
potential for electromagnetic
or other interference between
this equipment and other
devices. Should user
experience interference,
relocate this device or
minimize the use of the
affected equipment while this
device is in operation.
Summary of Contents for Cmax
Page 1: ......
Page 4: ......
Page 65: ......
Page 142: ...Page 3 of 14 ...
Page 148: ...Page 9 of 14 ...
Page 149: ...Page 10 of 14 ...
Page 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Page 164: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......