4-10
150830-685
Operator Manual
Operating Instructions
4.5.1 Auxiliary System
Operation
To operate system, proceed as follows:
1.
Flip pedal down (see Figure 4-5).
2.
Pull drawer out and remove auxiliary hand control.
3.
Return drawer to table.
4.
To operate table functions, pump pedal while pressing desired
function button on auxiliary hand control (see Figure 4-6).
NOTE: When using the auxiliary control system, note the following:
1) For
UNLOCK
and
SLIDE
functions, it is not necessary to pump
pedal.
2) Pressing any button on the auxiliary hand control deactivates
main hand control.
3) When any key is pressed on auxiliary hand control, the green
LED (see Figure 4-6) should light indicating system is active.
4) The ALS System is NOT active during auxiliary control
operation.
5) When table is in the auxiliary
UNLOCK
configuration, tabletop
functioning IS active.
LOCK
table before moving tabletop.
6) Reverse Orientation is NOT available in auxiliary control
operation. If patient was in reverse orientation when auxiliary
control was activated, change
TREND
to
REV. TREND
,
LEFT
TILT
to
RIGHT TILT
, etc.
WARNING – PERSONAL
INJURY AND/OR EQUIPMENT
DAMAGE HAZARD:
• ALS System is not active
during Auxiliary Control.
Ensure proper tabletop
movements to prevent
patient/staff injury and/or
equipment damage.
• During Auxiliary Control,
tabletop functions are
available even if table floor
locks are NOT engaged.
LOCK table before moving
tabletop or transferring
patient.
Figure 4-5. Auxiliary Control System
Summary of Contents for Cmax
Page 1: ......
Page 4: ......
Page 65: ......
Page 142: ...Page 3 of 14 ...
Page 148: ...Page 9 of 14 ...
Page 149: ...Page 10 of 14 ...
Page 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Page 164: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......