![Steris Cmax Operator'S Manual Download Page 30](http://html.mh-extra.com/html/steris/cmax/cmax_operators-manual_1374336030.webp)
3-4
150830-685
Operator Manual
Installation Instaructions
12.
Install leg section as follows:
NOTE: The Cmax Surgical Table is equipped with Hi-Lock™ locking
mechanism enabling quick and easy removal/installation of the leg
section with a safety lock feature.
a.
Using hand control, position table top seat section to
horizontal by pressing
LEG UP
button (refer to
S
ECTION
4.2,
H
AND
C
ONTROL
)
.
NOTE: Hand control LCD Display reads
LEG OUT
and displays a
blank box for Angle Value (see Figure 3-7).
b. Insert both leg section installation rods into seat frame until
fully engaged. Listen for "click" sound to indicate correct
installation (see Figure 3-8).
NOTE: The leg section should slide easily into the back frame (seat
section). Do not force.
c.
Pull leg section (both sides) to ensure correct installation.
d. Also, press
LEG UP
and
LEG DOWN
hand control buttons
and verify LCD Display reads a changing angle value to
ensure leg section has been installed properly.
This completes the Cmax Surgical Table installation. Refer to
for Operating Instructions
.
Figure 3-3. Connect Hand Control to Cmax Surgical Table
Control
Panel
Protective
Cover
Click
Summary of Contents for Cmax
Page 1: ......
Page 4: ......
Page 65: ......
Page 142: ...Page 3 of 14 ...
Page 148: ...Page 9 of 14 ...
Page 149: ...Page 10 of 14 ...
Page 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Page 164: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......