4-1
Operating Instructions
Operator Manual
150830-685
4.1 Operating Table
Operate the Cmax
™
Surgical Table as follows:
1.
Review Pinch Point Warnings as presented in
of this manual.
2.
Cmax Surgical Table is equipped with Intellipower
®
dual power
system. If table is to be operated on battery power, proceed to
Step 4
.
NOTE: For more information on the battery system, refer to S
ECTION
5.5, B
ATTERY
C
HARGING
P
ROCEDURE
.
3.
Verify power cord is plugged into both the facility outlet and table
receptacle and Main Power Switch is in ON position (lighted
green).
4.
For optimum performance, before attempting to operate table,
allow table to reach room temperature
(see Figure 4-1)
.
5.
After referencing
4.2.1, H
AND
C
ONTROL
F
UNCTION
B
UTTONS
D
ESCRIPTION
,
position tabletop to LEVEL.
NOTE: If the Hand Control AND Foot Control are installed, Hand
Control has priority. Also, the Cmax Table Head Section is manually
adjusted as follows (if Head Section is not installed, refer to S
,
Step 11
):
1) The head section may be positioned UP and DOWN 90°
from horizontal (level).
2) While holding head section securely in one hand, pull
release handle (spring loaded) located under right side of
head section (see Figure 4-2) and tilt head section to
desired angle (level in this case).
3) Release handle to lock head section at desired angle. Move
headrest slightly until ratchet mechanism locks headrest in
position.
WARNING – PERSONAL
INJURY HAZARD:
• Section 4.1, Steps 1 - 6 in
"Operating the Table" must
be executed before patient
transfer to the table.
Execution of Steps 1 - 6
with the patient on the table
could result in injury to
both the patient and the
operating room staff.
• Healthcare professionals
must ensure that patients
are positioned and
monitored so as to prevent
compromising respiration,
nerve pathways, or
circulation.
• Unanticipated table
movement could cause
patient injury. Patient must
be secured to the table in
accordance with
recommended positioning
practices.
• If the integrity of the
external protective earth
conductor installation or
arrangement is in doubt,
operate the table from its
internal power source.
WARNING – PERSONAL
INJURY AND/OR EQUIPMENT
DAMAGE HAZARD:
• Table is factory set to
operate at a certain voltage.
Use
of any other power
supply could result in
serious personal injury or
table damage.
• Use of this table in the
presence of flammable
anesthetics could result in
serious personal injury or
table damage.
Figure 4-1. Cmax Surgical Table
OPERATING INSTRUCTIONS
4
Summary of Contents for Cmax
Page 1: ......
Page 4: ......
Page 65: ......
Page 142: ...Page 3 of 14 ...
Page 148: ...Page 9 of 14 ...
Page 149: ...Page 10 of 14 ...
Page 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Page 164: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......