4-2
150830-685
Operator Manual
Operating Instructions
6.
Engage floor locks by pressing hand control
LOCK
button (see
7.
Before placing patient on table, ensure patient does not exceed
maximum patient weight limit (patient in normal orientation) of
1100 lb (499 kg) without patient posturing or tabletop slide
(except raise/lower); 1000 lb (454 kg) full patient posturing
(except tabletop slide);600 lb (272 kg) full patient posturing
(including tabletop slide). Assuming patient weight does not
exceed table weight capacity, transfer or position patient and
ensure patient is properly restrained (refer to
P
ATIENT
S
AFETY
S
TRAPS
,
in this manual).
NOTE: A patient grounding post/potential equalization terminal (male
connector, DIN 42801) is provided. The mating female connector for
patient grounding is not furnished by STERIS.
WARNING – PERSONAL
INJURY AND/OR EQUIPMENT
DAMAGE HAZARD: Placing a
patient whose weight exceeds
maximum patient weight limit
(patient in normal orientation)
of 1100 lb (499 kg) without
patient posturing or tabletop
slide (except raise/lower),
1000 lb (454 kg) with patient
posturing (except tabletop
slide), or 600 lb (272 kg) with
full patient posturing
(including tabletop slide)
could result in injury to the
patient and the operating
room staff, and serious
equipment damage.
WARNING – TIPPING
HAZARD:
• Do not place patient on the
table unless floor locks are
engaged.
• Do not use this table for
patients exceeding the
maximum patient weight
(patient in normal
orientation) of 1100 lb (499
kg) without patient
posturing or tabletop slide
(except raise / lower),
1000 lb (454 kg) with patient
posturing (except tabletop
slide), or 600 lb (272 kg)
with full patient posturing
(including tabletop slide).
• Center tabletop over
column before applying
chest compressions.
WARNING – PERSONAL
INJURY HAZARD: The Table
Leg Section is designed to
support the patient legs. Do
not put body weight on the
Leg Section.
Figure 4-2. Head Section Operation (Typical)
Handle
90°
90°
Summary of Contents for Cmax
Page 1: ......
Page 4: ......
Page 65: ......
Page 142: ...Page 3 of 14 ...
Page 148: ...Page 9 of 14 ...
Page 149: ...Page 10 of 14 ...
Page 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Page 164: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......