7-10
122993-622
Operator Manual
Troubleshooting
Table 7-2. Troubleshooting Chart - Alarms and/or Printouts (Cont'd)
PROBLEM
POSSIBLE CAUSE AND CORRECTION
9.
Display shows:
ALK. TANK
TOO LONG IN FILL
and printout message:
ALK.
TOO LONG IN FILL
Alkaline tank float did not signal full after tank fill time.
1.
Press
ALARM REPLY
touch pad to silence buzzer. Verify tank fill time
and alkaline tank full float.
2.
Press
CYCLE/START
touch pad to resume or
STOP/RESET
touch
pad to abort cycle.
10. Display shows:
RINSE TANK
TOO LONG IN FILL
and printout message:
RINSE TANK
TOO LONG IN FILL
11. Display shows:
ALK. TANK
FAILED TO REACH TEMP.
and printout message:
ALK. TANK
FAILED TO REACH TEMP
12. Display shows:
UNKNOWN
FAILURE DETECTED
and printout message:
UNKNOWN
FAILURE DETECTED
FAILURE NO. = XX
13. Display shows:
SERVICE DOOR
OPEN WHILE PROCESSING
and printout message:
SERVICE DOOR OPEN
WHILE PROCESSING
Rinse tank float did not signal full after tank fill time.
1.
Press
ALARM REPLY
touch pad to silence buzzer. Verify tank fill time
and rinse tank full float.
2.
Press
CYCLE/START
touch pad to resume or
STOP/RESET
touch
pad to abort cycle.
Alkaline Tank temperature did not reach set point within tank heat time.
Press
ALARM REPLY
touch pad to silence buzzer. Verify alkaline tank
steam system and tank heat time.
Alarm failure flag set to an unknown alarm.
1.
Press
ALARM REPLY
touch pad to silence alarm.
2.
Call STERIS.*
One of the service doors was opened while pumps were running.
1.
Press
ALARM REPLY
touch pad to silence buzzer. Verify all service
doors to see if any opened by accident.
2.
Verify all service door contacts.
3.
Close door and press to resume or press
STOP/RESET
touch pad to
abort.
* Service charges may be incurred. Consult your warranty for details.