3-7
Component Identification
Operator Manual
122993-622
A remote control column is installed on the unload side of the unit. This remote
control column is wired directly to the main control processor and includes a
control panel. Most washer functions can be directed from this control panel,
and the display screen concurrently shows the same cycle performance data
as the load-side control panel. There is no printer on remote control column.
Service doors are equipped with a safety switch to stop washer operation if one
of doors is opened during a cycle and to prevent start of washer operation if
door is not securely closed. Load-side and unload-side have a drive system
EMERGENCY STOP
pushbutton.
Load-side and unload-side control panels are also equipped with
STOP/
RESET
touch pads. Press touch pad once to stop cycle operation and twice
to abort cycle.
Additional controls affecting the drive system are located on the front of the
electrical control box (see Figure 3-7).
•
MANUAL DRIVE SYSTEM
- position switch to ON to run conveyor only.
•
ADJUSTABLE BELT SPEED
- position dial to desired conveyor speed from
low to high.
•
PROXIMITY SWITCH BYPASS
- position switch to ON to bypass proximity
switch.
3.5 Unload Side
Control Panel
(Double-Door Units)
3.6 Safety System
3.7 Drive System
Control
Figure 3-7. Drive Sytem Control
PROXIMITY SWITCH
BYPASS
ON
OFF
MANUAL DRIVE
SYSTEM
ADJUSTABLE BELT SPEED
LOW
HIGH