5-4
122993-622
Operator Manual
Cycle and Control Value Programming
13. Press
VALUE arrows
(up or down) to toggle between YES and NO.
Selecting NO by-passes acid wash tank function during cycle. Selecting
YES provides acid wash during cycle.
NOTE: Change Values mode may be exited at any time by pressing
SAVE
VALUES
touch pad. Control saves changed values and returns to selected
cycle menu.
14. Press
CHANGE VALUES
touch pad. Acid wash tank temperature guar-
antee option appears. Display shows:
GUARANTEE ACID TEMP?
NO
__ Indicates flashing position.
15. Press
VALUE arrows
(up or down) to toggle between YES and NO.
Selecting NO starts belt as soon as cycle starts. Selecting YES guarantees
acid wash tank temperature is equal to or greater than programmed set
point before belt starts.
NOTE: When changing values, note the following:
1) With any temperature guarantee feature enabled, the belt does not start
unless set point is reached.
2) Change Values mode may be exited at any time by pressing
SAVE
VALUES
touch pad. Control saves changed values and returns screen to
selected cycle menu.
16. Press
CHANGE VALUES
touch pad. Rinse tank temperature option
appears. Display shows:
CYCLE 1
SP=180.0 F
RINSE TANK
__Indicates flashing position.
17. To enter rinse temperature set point, press
CURSOR arrows
(left or right)
to select position and
VALUE arrows
(up or down) to select desired
number (0 to 9). Temperature set point is input as any number value to 1/
10
th
of a degree.
NOTE: When changing values, note the following:
1) With any tank temperature guarantee feature enabled, the belt does not
start unless set point is reached. It is important set point be an attainable value.
2) Change Values mode may be exited at any time by pressing
SAVE
VALUES
touch pad. Control saves changed values and returns screen to
selected cycle menu.
18. Press
CHANGE VALUES
touch pad. Rinse tank temperature guarantee
option appears. Display shows:
GUARANTEE RINS TEMP
NO
__ Indicates flashing position.
CHANGE
VALUES
CHANGE
VALUES
CHANGE
VALUES