6-11
Routine Maintenance
Operator Manual
122993-622
1.
Lock disconnect switch in
OFF
position and close building supply valves.
2.
Remove wires from each of two probes in wash and rinse recirculating
tanks.
NOTE: Make note of wire locations for reassembly. Reversing wires can
cause washer to malfunction.
3.
Lower water level in each tank below probe openings.
4.
Remove probes. Using a rag dampened with acid or descaler, remove
hard water deposits from each probe.
5.
Reinstall probes and attach wires in original positions.
6.
Re-energize washer utilities.
6.5.5 Clean Water Level
Sensing Probes
WARNING – ELECTRICAL
SHOCK AND/OR BURN
HAZARD: Disconnect all
utilities to washer before
servicing. Do not service the
washer unless all utilities
have been properly locked
out. Always follow local
electrical safety-related
work practice standards.
WARNING – BURN HAZARD:
Before performing any ser-
vice on the unit, wait until
chamber and piping cool to
room temperature.
WARNING – CHEMICAL
BURN AND/OR EYE INJURY
HAZARD: Detergents are
caustic and can cause ad-
verse effects to exposed
tissues. Do not get in eyes,
on skin, or attempt to swal-
low. Read and follow the
precautions and instruc-
tions on the detergent label
and in the Material Safety
Data Sheet (MSDS) prior to
handling the detergent, re-
filling the detergent con-
tainer, or servicing the de-
tergent injection pump.
Wear appropriate Personal
Protective Equipment
(PPE) whenever handling
the detergent or servicing
the detergent injection
pump and lines.