6-18
122993-622
Operator Manual
Routine Maintenance
The final rinse water temperature can be controlled manually by regulating the
flow of steam and hot water through the final rinse heat exchanger.
Adjust final rinse water temperature between 185 and 195 °F (85 and 91 °C)
as follows:
1.
If washer is running, press
STOP/RESET
touch pad.
2.
Completely close steam throttling valve and open fully final rinse water
throttling valve (see Figure 6-8).
3.
If necessary, allow solution/water in recirculating tanks to reach set
temperatures.
4.
Press
CYCLE START
touch pad.
5.
Verify reading on heat exchanger temperature gauge. Gauge indicates
temperature of final rinse water after water passes through heat ex-
changer.
6.
Slowly open steam throttling valve until temperature gauge reads desired
temperature.
7.
If gauge does not reach desired temperature after steam throttling valve
is fully open, slowly close final rinse water throttling valve until desired
temperature is reached.
IMPORTANT
: Final rinse water flow must be re-balanced if the final rinse
water throttling valve is adjusted (closed) while setting the final rinse water
temperature.
6.13 Final Rinse
Temperature
Controls
Figure 6-8. Final Rinse Temperature Controls
Final Rinse Water Throttle
Valve
Optional Dryer Steam
Throttle Valve
Heat Exchanger
Temperature Gauge
Steam Throttling Valve