4-3
Operating Instructions
Operator Manual
122993-622
The Basil
®
6000 Tunnel Cage Washer is equipped with a microcomputer
control capable of storing parameters for twelve distinct cycles. Authorized
operators have capability of customizing/programming all the cycles to meet
specific washing needs. For instructions on cycle programming or changing
cycle parameters, refer to
Section 5, Cycle and Control Value Programming.
On initial receipt of washer, each cycle is set with a basic demonstration cycle
consisting of sections of the unit, for example Pre-Wash, Alkaline Wash, Rinse,
Final Rinse, and Dry. Refer to
Table 4-1
for phase values of the demonstration
cycle.
To begin cycle operation:
1.
Set
POWER-OFF/STANDBY
switch, located behind printer door, to
POWER
.
Unit name temporarily appears, then display shows first cycle menu:
CYCLE 1
CYCLE 2
CYCLE 3
CYCLE 4
_ Indicates flashing position
and printer message is:
* CONTROL ON
HH:MM:SS
MM:DD:YY
===========================
MODEL 6XXX
S/N 36XXXXXXXXXX
============================
2.
Press
CYCLE MENU
touch pad until desired cycle menu appears. Display
shows:
CYCLE 9
CYCLE 10
CYCLE 11
CYCLE 12
_ Indicates flashing position
3.
Press
SELECT CYCLE
touch pad until desired cycle name flashes.
Display shows:
CYCLE 9
CYCLE 10
CYCLE 11
CYCLE 12
_ Indicates flashing position
4.
When desired cycle name is flashing, press
CYCLE/START
touch pad.
Name of selected cycle appears and remains displayed for five seconds.
Display shows:
CYCLE 9
CYCLE 10
CYCLE 11
CYCLE 12
__ Indicates flashing position
CYCLE
MENU
SELECT
CYCLE
CYCLE
START
Table 4-1. Demonstration
Cycle Phase Values
Cycle
Temperature
Phase
Pre-wash
Same as
Rinse Tank
Alk. Wash
140°F (60°C)
OR
Acid Wash
140°F (60°C)
Rinse
180°F (82°C)
Final Rinse
180°F (82°C)
Drying
Manually Adjusted
WARNING – ELECTRICAL
SHOCK AND/OR BURN
HAZARD: POWER-OFF/
STANDBY switch does not
shut ac power off to the
unit.
4.4 Cycle Operation