Chapter 15: Specifications
15-2
Operation Manual SurgiVet
®
V3404 Plus
Pulse Strength
Range:
0 - 15 bpm, indicates logarithmic strength of patients pulse. Not proportional to pulse
volume.
Display:
10 – segment graphics bargraph (update rate = 60Hz)
(Display is not proportional to pulse volume)
Pulse Rate
Range:
20 – 350 bpm
Accuracy:
> of ± 2% or ± 2 bpm (20 to 350 bpm)
Alarm Ranges:
Low 20-350 bpm (1 bpm steps) and Off
High 20-350 bpm (1 bpm steps) and Off
Display:
3-digit 0.43 inch high LED’s (update rate = 1 Hz)
Averaging:
Normal (8 seconds) or Slow (16 seconds) average (selectable)
ECG
Configuration:
3 or 5 lead
Lead Selection:
I, II, III (Standard Limb Lead Configuration),
aVR, aVL, aVF and V.
Gain:
X1, X2, X4, X8
Input Range:
-5.0mV to +5.0mV
DC Offset Voltage:
-300mV to +300mV
Frequency Response:
0.5Hz to 40Hz
Input Impedance:
> 5 Mohms differential, as required and tested by ANSI/AAMI EC-13
System Noise, max.:
30µV p-p
CMRR:
100db @ 50/60Hz
Leakage Current:
< 10µA
Patient Isolation:
> 4000VAC
Pace Detect:
Yes (minimum pulse amplitude of 3 mV)
Pace Reject:
Yes
Sweep Speed:
12.5, 25 or 50 mm/sec. (display update rate of 60 Hz)
Selectable 50Hz/60Hz
noise filter
Yes
Selectable QRS
threshold detection
Yes (0.10mV, 0.12mV,0.15mV, 0.40mV)
Heart Rate (ECG)
Range:
20 – 350 bpm
Accuracy:
> of ± 2% or ± 2 bpm (20 to 350 bpm)
Alarm Ranges:
Low 20-350 bpm (1 bpm steps) and Off
High 20-350 bpm (1 bpm steps) and Off
Display:
3-digit 0.43 inch high LED’s (update rate = 1 Hz)
Averaging:
8 seconds (not user selectable)