Chapter 15: Specifications
Operation Manual SurgiVet
®
V3404 Plus
15-3
Respiration
Range:
Sensitivity level 100, used for larger animals
Sensitivity level 10, used for small animals
Rate:
5 to 150 BrPM (breathes per minute)
Accuracy:
> of +/-2% or +/-2 BrPM over full range
Bandwidth:
0.16 to 2 Hz
Minimum signal:
10mL (Volume of air passing thermistor)
Display:
3-digit 10X14 pixel font (EL Panel) (update rate = 1Hz)
Splits time on screen with temperature, when on.
Temperature
Range:
0 to 60 °C (32 to 140 °F)
Accuracy:
±1 °C
Display:
3-digit 10X14 pixel font (EL Panel) (update rate = 1Hz)
Splits time on screen with respiration, when on.
Alarms:
No alarms for temperature
Alarms/Alerts/Indicators
Limit Exceeded or
Matched:
SpO
2
and Pulse Rate, associated display will flash ON and OFF, red ALARM LED will
flash, Audible will sound (if not silenced) LOW or HIGH and the parameter in violation
will be displayed in the message area.
ASYSTOLE Alarm:
ALARM LED will flash, Audible will sound (if not silenced), ASYSTOLE message on
graphics screen to flash
Audio Disabled:
SILENCED BELL icon (displayed on EL) indicates temporary or indefinite status
SpO2 Sensor:
SpO
2
Sensor message indicates an oximeter sensor problem, Audible will sound (if
not silenced), yellow ALERT LED continuously lit.
SpO2 Search:
SpO
2
SEARCH message indicates that device is searching for pulse
Artifact:
ARTIFACT message informs operator about excess of artifacts
ECG/Lead Fail:
ECG LEAD FAIL message indicates a lead off condition or ECG amplifier overload,
yellow ALERT LED is on continuously, Audible will sound (if not silenced)
Charge:
Green CHARGE LED indicates external power and battery is charging
Low Battery:
LOW BATTERY message indicates battery operation has approximately 10 minutes
remaining and should be charged, audible 5 beep burst will sound
Pace Detect:
A line will be drawn on the screen when a pace maker signal is detected.
Temper Sensor:
Temperature sensor not installed or not working
I/O’s
SpO
2
Sensor:
DB-9 connector with locking mechanism, compatible with all SurgiVet
®
oximeter
sensors
ECG Patient Cable:
AAMI 6-pin connector
Power:
standard DC power supplied by wall mount or table top unregulated AC adapter
Serial/Analog:
RS232C (DB-9 pin) serial output of data, 9600 baud, 8 bits. Also connects to optional
digital (remote alarm) interface. Two analog outputs can be configured to any of the
parameters monitored.