Chapter 5: ECG
5-2
Operation Manual SurgiVet
®
V3404 Plus
Using the ECG Parameter
If not familiar with this monitor and the ECG parameter, follow this chapter’s sections in order:
Setting Up ECG Monitoring
Ensure the ECG parameter is turned on. Refer to the
ECG
Menu for enabling the monitor.
Attaching the Patient
Attach the ECG leads and connect the ECG patient cable to the monitor.
ECG Display
Shows a typical screen and explains the ECG display’s features.
ECG Menu
Choose the ECG limb lead selection, size, pace detect, and number of leads
ECG Messages
Defines the ECG alarms, alerts, and messages.
Setting Up ECG Monitoring
Follow the steps in
Chapter 3: Setting Up the Monitor
. The remainder of this chapter assumes the monitor is
installed and turned on.
The ECG parameter can be turned on or off in the ECG menu. To ensure the ECG parameter is on, press the
b
key. Use the ARROW (
no
) keys to highlight the
ECG
option. Press the
b
key again to enter the ECG menu.
Use the ARROW (
no
) keys to highlight the
ECG Monitor
option, and confirm the option is set to
YES
.
Calibration Waveform
The calibration waveform can be activated to check the accuracy of the printer and/or analog output waveform
as well as the display. The calibration waveform is activated as follows:
When the monitor is turned on, simultaneously press the Wave/Trend
a
and Menu/Enter
b
keys to enter
the
Service Mode
.
Use the ARROW (
on
) keys to highlight the
Calibration
option.
Use the ARROW (
on
) keys to toggle between
ON
and
OFF
, then select desired option by pressing the
Menu/Enter
b
key.
1.
2.
3.