Chapter 11: Analog Output
Operation Manual SurgiVet
®
V3404 Plus
11-1
Chapter 11: Analog Output
Analog Out Menu
The
ANALOG OUT
menu allows the user to select parameter, waveform, or calibration data to output on each of
two analog output channels. All output data is scaled to the range [0-1Volt].
Figure 10.1: Analog Out Menu
To select the data to be output to any analog channel:
Select
ANALOG OUT
from the Main Menu.
Use the ARROW keys (
no
) to select the channel number
1
or
2
. Then press
b
.
Use the ARROW keys (
no
) to select the desired parameter (
ECG Waveform
,
Plethysmogram
,
%SpO2
,
Heart Rate
,
Resp Rate
,
Temperature
).
Press
b
to accept the value.
Select
[EXIT]
or press
a
to exit the menu.
1.
2.
3.
4.
5.
RR
/m
25
Q
CHANNEL PARAMETER TO
NUMBER OUTPUT
1 ECG WAVEFORM
2 0 V
[EXIT] (analog out menu)