Chapter 3: Setting Up the Monitor
Operation Manual SurgiVet
®
V3404 Plus
3-3
Working With System-Wide Settings
This section describes working with system-wide settings using the
SETUP/VOLUME
menu.
Setup/Volume Menu
The following system-wide settings are viewed and/or adjusted from the Setup/Volume menu:
Figure 3.1: Setup Menu
SETUP
The waveform sweep speed can be set for fast (25 mm/sec) and slow (12 mm/sec) speeds. The heart rate source
can be set to either Oximeter or ECG.
VOLUME
Indicates the Alarm and Pulse volume. Allows the volume to be adjusted in the range of 1 to 15. Pulse volume
can be set to OFF. Alarm volume can not be shut off.
Turning Parameter Monitoring On and Off
Monitoring for the following parameters can be turned on or off:
ECG (ECG waveform, heart rate)
NOTE! Disabling the ECG parameter also disables Respiration.
Respiration (RR/M)
Oximeter (Plethysmogram, SpO
2
%, pulse rate, pulse strength)
Temperature (°C / °F)
When turned off, displays, indicators, and alarms related to the parameter are disabled. The parameter occupies
a space on the display and in the serial output, but its value is shown as dashes (---). If the parameter is assigned
to an analog channel, that channel shows 0 Volts. If a parameter has a waveform, its waveform area shows a
monitor off message.
To turn a parameter’s monitoring on or off, do the following:
From the Main menu, select the desired parameter’s menu item:
ECG
,
OXIMETER
,
RESPIRATION
, or
TEMPERATURE
.
In the parameter’s menu, select the monitor
on/off
item.
Press
b
and use the ARROWS (
no
) to adjust the setting.
Press the
b
key to accept the setting.
Select
[EXIT]
or press
a
to exit menus.
•
•
•
•
1.
2.
3.
4.
5.
RR
/m
25
Q
Sp02 SENSOR!
SETUP
SWEEP SPEED 12.5mm/sec
BPM SOURCE OXIMETER
VOLUME
ALARMS
PULSE
[EXIT] (setup/volume menu)