Appendix A: Guidance and Manufacturer’s Declaration
Operation Manual SurgiVet
®
V3404 Plus
A-1
Appendix A: Guidance and Manufacturer’s Declaration
Guidance and Manufacturer’s Declaration
The V3404 Plus is intended for use in the electromagnetic environment specified in the tables within this
appendix.
NOTE! The customer or user of the V3404 Plus should ensure that it is used in such an environment.
Electromagnetic Emissions - Emissions Test
EMISSIONS TEST
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT GUIDANCE
RF emissions
CISPR 11
Group 1
The V3404 Plus uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The V3404 Plus is suitable for use in all establishments, including:
Domestic establishments.
Establishments directly connected to the public low-
voltage power supply network that supplies buildings used
for domestic purposes.
•
•
Harmonic emissions
IEC 61000-3-2
NA
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
NA
Electromagnetic Emissions – Immunity
IMMUNITY
ELECTROMAGNETIC ENVIRONMENT GUIDANCE
Electrostatic discharge
(ESD)
IEC 61000-4-2
IEC 60601 TEST LEVEL Floors should be made of:
Wood
Concrete
Ceramic tile
If floors are covered with synthetic material, the relative
humidity should be at least 30%.
•
•
•
± 6 kV contact
± 8 kV air
•
•
COMPLIANCE LEVEL
± 2 kV contact
± 8 kV air
•
•