Chapter 15: Specifications
15-4
Operation Manual SurgiVet
®
V3404 Plus
Optional Printer and Serial Interface
A graphics printer output is available. It runs in either of the following modes, and is controlled by the Start/Stop
key on the printer panel.
The connection will be a serial one with the following settings: 9600bps, No parity, 8 data bits, 1 stop bit.
Data Log Mode:
When this mode is set through menu, every 5 sec data line is sent out. Data line
includes SpO
2
, Pulse Rate or Heart Rate, Pulse and Respiration (if installed).
Trend Printing:
When this mode is set through menu, trend data is sent with a speed defined by the
CTS return line (defined by printer or computer).
Waveform Printing:
The user will be able to print a real time plot of ECG and/or a Plethysmogram.
Generic Printing:
There will be the option to send text data to any RS-232 port.
Trend Storage
Parameters:
SpO
2
, Pulse Rate (Heart Rate), Signal Strength, Respiration and Temperature
Interval:
30 sec non-adjustable
Capacity:
12 hours of storage
Sound Control
Pulse Beep Volume:
15 levels of volume and off
Alarm Beep Volume:
15 levels of volume and silenced
Sound level :
less than 85dBA.
Brightness Control
None
Analog Outputs
Channels:
Two, configurable through menu
Range:
0 to 1 VDC
Differential Accuracy :
± 10mV
Output Offset:
+40/-0 mV
Max. Load:
3k Ohm, 500pF
Digital Alarm Output to Remote Sensing Location - Nurse Call
Configuration:
Optional, attached and powered by communication port
Output:
DC switch (MOSFET)
Max. voltage:
24 VDC
Max. load current:
0.5 A continuous, 5 A peak
Max. On resistance:
0.15 Ohm
Power
Battery:
1 to 1.5 hours of continuous operation max. charge time 6 hours.
External:
12VDC @ 1A provided by external wall mount transformer with 4kV isolation